Jump back to chapter selection.


Table of Contents

6.1 Lasers
6.2 Frequency Tuning
6.3 Phase-Sensitive SHG


6 Experimental Methods of Nonlinear Optics

As already mentioned in the previous chapter, observing nonlinear effects requires high intensity fields. This, in turn, requires high intensity light sources to compensate for the difference in the first and second order susceptibility χ(1) and χ(2), the respective susceptibility tensors in the Taylor expansion of Pi(ω). In the following, three light sources are compared regarding their intensity:

Light source Peak intensity
Sun 0.1 W/cm2
Arc lamp 100 kW/cm2
Pulsed laser 1061013 W/cm2

6.1 Lasers

As becomes clear from the previous table, the laser is the clear winner for use in nonlinear effects. Additionally, it offers high coherence, a well-defined and narrow spectral band, as well as a well-defined polarisation state. These are important and desirable properties when designing an experiment. Two common light sources used in nonlinear optics are neodymium-doped yttrium aluminium garnet lasers (Nd:YAG laser) or titanium-sapphire lasers (Ti:Sa laser). Some output parameters are presented in the next table:

Laser λ Δt Repetition rate Energy Application
Nd:YAG 1064 nm 5 ns 10100 Hz 10100 mJ Transmission experiments on bulk samples
Ti:Sa 800 nm 100 fs 1 kHz 110 mJ Reflection setups

6.2 Frequency Tuning

A laser initially only emits at its own characteristic, well-defined frequency governed by the lasing transition. However, for experimental purposes, it is often required for the frequency to be tunable to some degree. This can be done by exploiting nonlinear processes; commonly used methods include parametric conversion and harmonic generation.

Parametric Conversion

Parametric conversion is a process in which an intense beam of light (the pump beam) is sent into a nonlinear crystal, emitting two beams at lower frequencies. These are usually labelled as the signal beam S and the idler beam I. One can think of the pump photons being 'split' into pairs of photons, conserving energy and momentum. These conservation laws lead to two conditions:

Energy conservation:{ωp=ωS+ωIωp=ωS+ωIMomentum conservation (collinear):{kp=kS+kInpωp=nSωS+nIωI

Since the refractive indices change with the orientation of the crystal relative to the light propagation direction and polarisation (birefringence), the crystal alignment is used for continuous frequency tuning.

Harmonic Generation

Harmonic generation is the process of generating higher harmonics (2ω,3ω,) of the fundamental frequency ω in nonlinear crystals. Furthermore, difference frequency generation may also be used via ω1,ω2ω1ω2.

Both methods combined may then be applied to the two laser systems introduced above. As an example, consider third-harmonic generation (THG) from an Nd:YAG laser (1064 nm), so 1064 nmTHG355 nm. Parametric down-conversion then allows continuous access to two ranges: 400710 nm and 7103000 nm. Similarly, a titanium-sapphire laser system may reach the range 21020000 nm via parametric conversion and multistep frequency mixing. Typical conversion efficiencies reach 2030% in a common material such as BBO (βBaB2O4).


6.3 Phase-Sensitive SHG

In a standard SHG experiment, the signal is measured as an intensity (for example, via a photodiode), such that all phase information is lost:

PSHG=ε0χ(2)EEISHG|PSHG|2.

Then, consider as an example again the spin-reversal domains discussed earlier. Here, the time-reversal operator T^ transforms two 180 domains into one another:

T^(domain+)=(domain).

Applying this operator to the polarisation then results in

T^P(2ω)T^χNLe=!χNLeP(2ω).

This then reveals that the polarisation generated in the two domains differs by a phase of 1=eiπ. This corresponds to a π phase shift in the polarisation vector. However, as already mentioned, this phase information is lost when measuring only the intensity. Thus, in order to resolve the different domains, it is not enough to extract the intensity information alone; one additionally has to measure the phase information. An interference measurement with a reference wave (such as in holography) allows the full information to be obtained:

ISHGtotal(domain+)|PSHG+(2ω)+Pref(2ω)|2=|P++Pref|2,ISHGtotal(domain)|PSHG(2ω)+Pref(2ω)|2=|P++Pref|2.

The only reason the antiferromagnetic domains could be observed in Cr2O3 using SHG is that this material intrinsically provides an i-type reference wave via the magnetic dipole (MD-SHG) contribution. Very often, however, no intrinsic reference is present, and a reference needs to be generated externally. By passing the fundamental beam through a quartz plate (or another suitable material) after it has passed through the sample, the fundamental wave generates an SHG reference wave. Interference then allows the extraction of phase as well as amplitude of the signal from the sample. The experiment is schematically depicted in the following figure:

Attachments/Script (Unofficial) 12.webp|700

Such a setup requires the reference crystal (and the sample) to be transparent at both the fundamental and the harmonic frequency, and the reference crystal additionally has to have a high conversion efficiency. Optical quartz plates usually fulfil these requirements while being available at good quality.

The experiment has two degrees of freedom to adjust:

In general, the relative phase should be adjusted such that ϕrel =0 or ϕrel =π (corresponding to 0 or 180), for constructive or destructive interference, respectively. The amplitudes should then be chosen to be equal, Aref =ASA, such that the contrast is maximal:

ISHGmin=0 and ISHGmax=|2A|2=4A2.

Usually, the phase is continuously adjusted by a Soleil-Babinet compensator. The alignment of the reference crystal is chosen such that the polarisation of the reference beam is orthogonal to the signal beam, and the Soleil-Babinet compensator is aligned such that these two polarisations coincide with its ordinary and extraordinary axes. The birefringence of the compensator then allows tuning of the relative phase. Finally, a linear polariser is used to adjust the relative amplitudes. By rotating the polariser, the projection of one amplitude onto the other is varied. The following figure shows the projection of the magnetic signal onto the reference signal. By adjusting the rotation of the polariser, the relative amplitude may be adjusted:

Attachments/Script (Unofficial) 13.webp|700

As became clear, a phase-sensitive SHG measurement requires good interference. Therefore, divergence and propagation effects like beam broadening between the sample and reference crystal can lead to the loss of spatial coherence. Temporal coherence, indicated by the coherence time and length, is related to the spectral width of the light. For a given spectral lineshape, the product of spectral width Δf and coherence time τc is approximately constant. For instance, for a Gaussian spectrum, Δfτc0.441 (using FWHM definitions).

The interference signal intensity can be written as

I(φ)=Isig+Iref+2αIsigIrefcosφ,

where Isig is the intensity from the sample SHG, Iref is the reference SHG intensity, φ is the relative phase between the two beams, and α is the degree of coherence (or fringe visibility). For perfect coherence, α=1. The degree of coherence α can be expressed using the normalised auto-correlation function Γ11(τ) and the spectral density σ(ν):

α(τ)=Γ11(τ)Γ11(0)=0σ(ν)e2πiντ dν0σ(ν)dν={eτ22τc2Gauss (field)e2|τ|τcLorentz (field)

where τ is the delay between the interfering beams and τc is the coherence time.

This translates into a fringe contrast C:

C=ImaxIminImax+Imin=α2IsigIrefIsig+Iref,

where Imax=max[I(φ)] and Imin=min[I(φ)]. This yields:

C{=αif Isig=Iref<αotherwise 

However, loss in contrast is not only due to the loss of temporal coherence but also because the spatial overlap and wavefront matching of the sample and reference beams may not be perfect due to propagation effects. Achromatic beam imaging, as shown in the following figure, can help to overcome problems of spatial decoherence:

Attachments/Script (Unofficial) 14.webp|700

Spatial decoherence due to effects like beam broadening can be mitigated by careful optical design, for instance using appropriate imaging optics like a spherical mirror. Temporal decoherence is minimised by using narrowband light sources with a correspondingly large coherence time.